AMPLIFIER AIDED CURRENT TRANSFORMER
MODEL CT5001

Specifications at 25 degrees C:
 
Current Ratio 
5000:1 Amperes
Transformation Accuracy
0.01% + 0.001% of full scale
Full Scale Input Current 
5000 Amperes
Primary Burden 
Resistance of Primary Circuit
Absolute Maximum Input 
5500 Amperes
Maximum Secondary Current 
1.1 Ampere
Maximum Secondary Voltage
         @ 1 ohm burden 
1.1 volts
         @ 4 ohms or greater
2 volts
Frequency Range
40 – 1000 Hz.
Phase Shift 
Less than 0.006 degrees
Distortion
Less than 0.1%
AC Calibration Cycle 
None Required
AC Temperature Coefficient 
5 ppm/degree C
DC Offset 
Less than 10 microamperes
DC Temperature Coefficient
10 ppm
DC Calibration Cycle 
3 Months
Transient response time
Greater than 20 seconds
Operating Temperature 
0 – 40 degrees C
Relative Humidity 
5 – 90% non condensing
Power Consumption 
120 VA’s maximum
Power Requirement – as marked 
115 or 230 V +/-5% 60 Hz. 
Weight 
12 pounds
Physical Size
    Amplifier 
7”x3”x5” WxHtxD (Enclosure only)
    Sense Transformer
3.5"I.D.x8.5"O.D.x2.25"Ht.
    Connecting Cable 
24 inches